{ "schema_version": "1.4.0", "id": "GHSA-64pp-r5x2-fgff", "modified": "2022-05-13T01:46:01Z", "published": "2022-05-13T01:46:01Z", "aliases": [ "CVE-2017-5149" ], "details": "An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.", "severity": [ { "type": "CVSS_V3", "score": "CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H" } ], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2017-5149" }, { "type": "WEB", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A" }, { "type": "WEB", "url": "http://www.securityfocus.com/bid/95331" } ], "database_specific": { "cwe_ids": [ "CWE-476" ], "severity": "HIGH", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2017-02-13T22:59:00Z" } }