{ "schema_version": "1.4.0", "id": "GHSA-82mj-h4mj-5q4v", "modified": "2022-05-13T01:38:54Z", "published": "2022-05-13T01:38:54Z", "aliases": [ "CVE-2015-1012" ], "details": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.", "severity": [ { "type": "CVSS_V3", "score": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N" } ], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2015-1012" }, { "type": "WEB", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01" } ], "database_specific": { "cwe_ids": [ "CWE-200" ], "severity": "HIGH", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2019-03-25T19:29:00Z" } }