{ "schema_version": "1.4.0", "id": "GHSA-85xx-4vmh-93wq", "modified": "2022-05-13T01:38:53Z", "published": "2022-05-13T01:38:53Z", "aliases": [ "CVE-2015-3952" ], "details": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.", "severity": [ { "type": "CVSS_V3", "score": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N" } ], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2015-3952" }, { "type": "WEB", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" } ], "database_specific": { "cwe_ids": [ "CWE-200" ], "severity": "HIGH", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2019-03-25T16:29:00Z" } }