{ "schema_version": "1.4.0", "id": "GHSA-cv4q-4xgp-cr7m", "modified": "2025-05-22T18:31:09Z", "published": "2022-05-13T01:34:29Z", "aliases": [ "CVE-2018-14781" ], "details": "Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the \"easy bolus\" and \"remote bolus\" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.", "severity": [ { "type": "CVSS_V3", "score": "CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N" } ], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2018-14781" }, { "type": "WEB", "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html" }, { "type": "WEB", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02" }, { "type": "WEB", "url": "http://www.securityfocus.com/bid/105044" } ], "database_specific": { "cwe_ids": [ "CWE-287", "CWE-294" ], "severity": "MODERATE", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2018-08-13T21:48:00Z" } }