{ "schema_version": "1.4.0", "id": "GHSA-jfh4-jcmh-hq55", "modified": "2022-05-24T19:09:36Z", "published": "2022-05-24T19:09:36Z", "aliases": [ "CVE-2021-37165" ], "details": "A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMsg function through the pendingQ, and may lead to remote code execution.", "severity": [], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2021-37165" }, { "type": "WEB", "url": "https://www.armis.com/PwnedPiper" }, { "type": "WEB", "url": "https://www.swisslog-healthcare.com" }, { "type": "WEB", "url": "https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37165-bulletin---overflow-in-hmiprocessmsg.pdf?rev=2e2678dab62b41ba999cd6d1e03974ca&hash=F465ACE2C7FAED826B52FE996E36ACEC" }, { "type": "WEB", "url": "https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20,%20%20CVE-2021-37164%20%204%20more%20rows%20" } ], "database_specific": { "cwe_ids": [ "CWE-120" ], "severity": "CRITICAL", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2021-08-02T11:15:00Z" } }