{ "schema_version": "1.4.0", "id": "GHSA-7fpg-565m-qfcv", "modified": "2022-07-13T00:01:26Z", "published": "2022-05-24T19:12:12Z", "aliases": [ "CVE-2021-33886" ], "details": "An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.", "severity": [ { "type": "CVSS_V3", "score": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H" } ], "affected": [], "references": [ { "type": "ADVISORY", "url": "https://nvd.nist.gov/vuln/detail/CVE-2021-33886" }, { "type": "WEB", "url": "https://www.bbraunusa.com/en.htm" }, { "type": "WEB", "url": "https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump" } ], "database_specific": { "cwe_ids": [ "CWE-20" ], "severity": "HIGH", "github_reviewed": false, "github_reviewed_at": null, "nvd_published_at": "2021-08-25T12:15:00Z" } }